We use cookies in order to guarantee the best possible service. If you continue browsing the site you consent to our cookie use.

Configurations:

PicoScope9301-209301-309302-209311-209321-209341-209341-30
20 GHz sampling oscilloscope       X X X  
30 GHz sampling oscilloscope X X X       X
2 channels X X X X X    
4 channels           X X
Clock recovery (11.3 Gb/s)     X   X    
Optical input (9.5 GHz)         X    
Integrated TDR/TDT (60 ps / 2.5 to 7 V)       X      

30 GHz bandwidth in a compact USB instrument

PicoScope 9300 Series sampling oscilloscopes use triggered sequential sampling to capture high-bandwidth repetitive or clock-derived signals without the expense or jitter of a very high-speed clocked sampling system such as a real-time oscilloscope. 25 GHz bandwidth allows measurement of 14 ps transitions, and low sampling jitter enables timing resolution down to 0.064 ps. Sequential sampling rate of 1 MS/s, unsurpassed by other sampling oscilloscopes, enables rapid building of waveforms, eye diagrams and histograms.

These two and four channel units occupy very little space on a workbench and are small enough to carry with a laptop for on-site testing. Furthermore, instead of using remote probe heads attached to a large bench-top unit, you can position the PicoScope 9300 right next to the device under test and connect to it with short, low-loss coaxial cables.

Everything you need is built into the oscilloscope, with no expensive hardware or software add-ons to worry about. Alternatively, you can use your PicoScope 9300 with a stand-alone PG900 TDR/TDT differential fast pulse generator to gain the extra versatility and configurability of independent high-performance source and measurement instruments.

Watch video - PicoScope 9300 introduction

 

Trigger modes

2.5 GHz direct and up to 15 GHz prescaled trigger

Sampling oscilloscopes accept their trigger from a separate input, either directly for repetition rates up to 2.5 GHz or via a prescaling divider input, for repetition rates up to 15 GHz (14 GHz on 15 and 20 GHz models).

Built-in 11.3 Gb/s clock data recovery trigger

To support serial data applications in which the data clock is not available as a trigger, or for which trigger jitter needs to be reduced, the PicoScope 9302 and 9321 include a clock recovery module. This continuously regenerates the data clock from the incoming serial data or trigger signal and can do so with reduced jitter even over very long trigger delays or for pattern lock applications. A divider accessory kit is included to route the signal to both the clock recovery and oscilloscope inputs.

 

9.5 GHz optical model

The PicoScope 9321-20 includes a built-in precision optical-to-electrical
converter. With the converter output routed to one of the scope inputs
(optionally through an SMA pulse shaping filter), the PicoScope 9321-20 can analyze standard optical communications signals such as OC48/STM16, 4.250 Gb/s Fibre Channel and 2xGB Ethernet. The scope can perform eyediagram measurements with automatic measurement of optical parameters including extinction ratio, S/N ratio, eye height and eye width. With its integrated clock recovery module, the scope is usable to 11.3 Gb/s.

The converter input accepts both single-mode (SM) and multi-mode (MM)
fibers and has a wavelength range of 750 to 1650 nm.

 

TDR/TDT analysis

The PicoScope 9311 oscilloscopes feature built-in step generators for time-domain reflectometry and transmission measurements. The 9311-15 integrates a single rising step generator suited to single-ended TDR/TDT applications, while the 9311‑20 features deskewable rising and falling step generators suited to single-ended and differential measurements. These features can be used to characterize transmission lines, printed circuit traces, connectors and cables with 16 mm resolution for impedance measurements and 4 mm resolution for fault detection.

 

tdr connections

The PicoScope 9311-15 and 9311-20 generate 2.5 to 7 V steps with 60 ps rise time from built-in step recovery diodes. They are supplied with a comprehensive set of calibrated accessories to support your TDR/TDT measurements, including cables, signal dividers, adaptors, attenuator and reference load and short.

The PicoScope 9311-20 TDR/TDT model includes source deskew with 1 ps resolution and comprehensive calibration, reference plane and measurement functions. Voltage, impedance or reflection coefficient (ρ) can be plotted against time or distance.

An alternative approach to TDR/TDT capability is to pair any 9300 Series scope with a standalone PG900 pulse generator. These instruments include similar differential step recovery diode step generators and also offer an option of 40 ps tunnel diode step generation. This brings extra flexibility and the ability to remotely position the pulse source. The generators also enable TDT and TDR with the PicoScope 9301, 9302 clock recovery, 9321 optical and 9341 4-channel sampling oscilloscopes.

 

Built-in signal generator

All the PicoScope 9300 Series scopes can generate industry-standard and custom signals including clock, pulse and pseudo-random binary sequence. You can use these to test the instrument’s inputs, experiment with its features and verify complex setups such as mask tests. AUX OUTPUT can also be configured as a trigger output.